Fluke 2620A Benutzerhandbuch

Seite von 197
Additional Considerations
When Measuring Resistance or Temperature (Rtd)
5
5-7
12
11
13 14 15 16 17 18 19 20
2
1
3
4
5
6
7
8
9
10
SOURCE
(4-WIRE)
SENSE
(4-WIRE)
H  L
H  L
H  L
H  L
H  L
H  L
H  L H  L
H  L H  L
H  L
H  L
H  L
H  L
H  L
H  L
H  L H  L
H  L H  L
2-WIRE (2T) CONNECTION
RESISTANCE
OR
RTD SOURCE
USE H AND L TERMINALS FOR ANY CHANNEL.
•   CHANNEL 0 ON FRONT PANEL
•   CHANNELS 1 THROUGH 20 ON REAR 
     PANEL INPUT MODULE (CHANNEL 8 SHOWN HERE).
12
11
13 14 15 16 17 18 19 20
2
1
3
4
5
6
7
8
9
10
SOURCE
(4-WIRE)
SENSE
(4-WIRE)
H  L
H  L H  L H  L H  L H  L
H  L H  L
H  L H  L
H  L
H  L H  L H  L H  L H  L
H  L H  L
H  L H  L
4-WIRE (4T) CONNECTION
RESISTANCE
OR
RTD SOURCE
USE H AND L TERMINALS FOR TWO CHANNELS ON REAR PANEL INPUT MODULE. 
CONNECTIONS FOR CHANNEL 8 SHOWN HERE WITH CHANNEL 18 PROVIDING 
ADDITIONAL TWO CONNECTIONS.
FOR EACH 4-WIRE CONNECTION, ONE SENSE CHANNEL (1 THROUGH 10) AND 
ONE SOURCE CHANNEL (SENSE CHANNEL NUMBER +10 = 11 THROUGH 20) ARE USED.
oo16f.eps
Figure 5-2. 2T and 4T-Connections