Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. The most common cause of measurement error 
is bad connections.
 The following procedures illustrate how to make good connections.
CAUTION
Never mate a 50 ohm connector with a 75 ohm connector. The larger center 
pin of a male 50 ohm connector will destroy the contact fingers of a female 
75 ohm connector.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, 
conductive table mat. Refer to 
 for ESD 
precautions.
2. Visually inspect the connectors. Refer to 
.
3. If necessary, clean the connectors. Refer to 
.
4. Carefully align the connectors. The male connector center pin must slip concentrically 
into the contact finger of the female connector.
5. Push the connectors straight together.
CAUTION
Do not turn the device body. Only turn the connector nut. Damage to the 
center conductor can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is 
usually a slight resistance.
6. The preliminary connection is tight enough when the mating plane surfaces make 
uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both 
mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
7. Make sure the connectors are properly supported. Relieve any side pressure on the 
connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
1. Use a torque wrench to make a final connection. 
 provides information about 
the torque wrench recommended for use with this calibration kit. A torque wrench is 
not included in the calibration kit. Refer to 
 for part number and ordering 
information.