Troubleshooting
6
Troubleshooting
Alarms will alert you to any problems with your device. The instructions 
in this section will help you to identify the cause of the problem. If the 
fault persists or cannot be identified, do not try to open the device, and 
contact your HME provider.
An alarm will activate an audible signal. If you know which alarm has 
been triggered, consult the table below, otherwise contact your HME 
provider.
Table 3: Table of alarms (not an exhaustive list).
Alarm
Cause 
Solution
Mains alarm
The mains power cable 
has been disconnected.
Reconnect the mains 
power cord.
External power 
supply alarm
The external power 
supply cable has been 
disconnected. This alarm 
can only be triggered if 
an external battery is 
connected to your 
device.
Reconnect the external 
power supply cable. 
Low battery alarm
The charge level of the 
built-in battery is low 
(15% or less).
Connect the device to a 
mains power supply 
without delay to 
recharge the built-in 
battery.
Flat Battery alarm
The built-in battery is flat 
(charge level less than 
5%).
Recharge the built-in 
battery 
immediately by 
connecting the device to 
the mains supply.
Disconnection 
alarm
A component in your 
circuit is wrongly 
connected or 
disconnected.
Reconnect the circuit.