3.6.3  Installing devices in sandwich design
•  Push the first bolt 3 through the hole 5 of both flanges.
•  Screw on the nuts and washers to both ends of the bolt 3 but do not tighten them.
•  Install the second bolt through the holes 4.
•  Place the measuring sensor 1 between the two flanges.
•  Insert the gaskets 6 between measuring sensor 1 and flanges and align them.
•  Check that the flange is concentric.
•  Install the remaining bolts, washers and nuts. Do not yet tighten the nuts.
•  Turn the centring ring 2 in a counter-clockwise direction and align the device.
•  Check that the gaskets 6 are concentric; they must not protrude into the pipe cross-section.
•  Now tighten all nuts bit by bit alternately across the diagonal.
Figure 3-13: Installation using centering ring
1  Measuring sensor
2  Centering ring
3  Bolts with fixing nuts
4  Hole
5  Hole
6  Gasket