Customizing set-up parameters in DMS 
          Dosimeter setting 
053-644, Rev E  
Page   25 
 
eg4 & eg3 
 
25 
Dosimeter settings for the Edge 
Name Field 
*Note:  refer to Glossary for more detailed information.   
Heartbeat enable  
When enabled (or checked), the LED will flash green while running 
and flash amber when stopped.  
Response 
Either Slow Time Response or Fast Time Response is applied to 
the measurements.   (Typically Slow time response is used in the 
US market while Fast is used in the EU communities.)  
C-A enable 
When enabled (or checked), C minus A will be calculated. 
 Projected Time 
A selectable time parameter via the up/down arrows or type in a 
value.  The projected time value is used to make projected dose 
calculation for the dose measurement.   
Peak Weighting 
Select either A, C or Z for the peak frequency weighting.   
Dose indicator LED 
When enabled (or checked), the LED will flash red when the dose 
value exceeds the designated entered value.   
Setup name 
(Optional field)  A setup name may be entered into the Meter 1 
and/or Meter 2 fields.  When set, it appears on the top of  Edge’s 
display (example setup names:  OHSHA, MSHA, EU directive). 
Threshold enable 
When the checkbox is selected and a value is entered for meter 1 
and/or meter 2, the threshold is used to calculate Lavg, TWA, and 
Dose measurements.  
Upper Limit 
Select a UL dB level by the up/down arrows or by typing in a value.  
(Note:  The upper limit (UL) is a measurement setting in which the 
user selects a specific decibel level. The UL value records the 
amount of time that the Sound level is at or greater than the 
entered level.)  
RMS Weighting 
Select either A or C RMS Weighting.  (Note:  The RMS (root mean 
square) are the results from the Edge with the response time and 
weighting settings taken into account.)  
Criterion Level 
Select a Criterion level via the up/down arrow or by typing in a 
value.  (Note:  Criterion level is the average SPL that will result in 
100% dose over the criterion time and is typically set by a 
regulatory agency such as OSHA, EU directive.)  
Exchange Rate 
Enter in (or select with up/down arrows) 3, 4, or 5 dB.  ER refers to 
how sound energy is averaged over time. (OSHA uses a 5 dB 
doubling rate while most of the world uses 3 dB.) 
 
Table 1-2:  Dosimeter fields explained for the Edge