 12   Verification
 Table 2-1.  Test Equipment Required
 
 Type
 Required Characteristics
 
 Recommended Model
 Digital Voltmeter
1
 Resolution: 10 nV @ 1V
 Agilent 3458A
 
 Readout: 8 1/2 digits
 
 
 Accuracy: 20 ppm
 
 Current Monitor Resistor
1
 Agilent 6682A, 6683A, Agilent 6684A:
 Guildline 9230/300
 
   0.001
Ω ± 0.04%, 300A, 100W
 
 
 
 
 
 Agilent 6680A, 6681A
 Burster Type 1280
 
   100 
µΩ ± 0.05%, 1000A
 
 DC Power Supply
 DC Power Source with current capability
equal to UUT
 Agilent 6680A
 Electronic Load
 
 
 
 Resistor Load
 Range: Voltage and current range must
exceed that of supply under test. Power:
5.4KW minimum
 or
 5 Kilowatt minimum
 Agilent 6680A = 5.7 milliohms 4375W
Agilent 6681A = 13.8 milliohms 4640W
Agilent 6682A = 87.5 milliohms. 5040W
 Agilent 6683A = 200 milliohms 5120W
 Agilent 6684A = 312 milliohms 5120W
 3 each Agilent 6050A, w/3 each
Agilent 60504B per Agilent 6050A
for all units
 
 Oscilloscope
 Sensitivity: 1mV
 Agilent 54504A
 
 Bandwidth Limit: 20MHz
 
 
 Probe: 1:1 with RF tip
 
 
 RMS Voltmeter
 True RMS Bandwidth: 20MHz
 Agilent 3400B
 
 Sensitivity: 100 
µV
 
 
 Current Transformer
 0.1Volt per ampere: 1Hz to 20MHz
 Pearson Model 411
 Variable-Voltage
Transformer (AC Source)
 Power: 3 Phase 24KVA; Range:
180-235V 47 - 63Hz; 360- 440V
 47 - 63Hz
 Superior Powerstat
 1156DT-3Y, 0-280V, 50A,
 24.2 KVA or equivalent .
 
 GPIB Controller
2
 Full GPIB capabilities
 
 HP Series 200/300
 
1
  Required for Operation Verification Tests.
 
2  
Required for remote testing of 668xA models.
 
 
Electronic Load
  Many of the test procedures require the use of a variable load capable of dissipating the required power. If a variable
resistor is used, switches must be used for connecting, disconnecting, and shorting the load resistor. For most tests, an
electronic load (see Table 2-1) is easier to use than a variable resistor. However, an electronic load may not be fast enough
for testing transient recovery time or may be too noisy for testing noise (PARD). In these cases, fixed load resistors of
suitable power dissipation can be used with minor changes to the test procedures given in this chapter.