Agilent Technologies 6684A Benutzerhandbuch

Seite von 156
Backdating   149
A
Backdating
Manual backdating describes changes that must be made to this manual for power supplies whose serial numbers are lower
than those listed in the title page to this manual.
Look in the following table and locate your Agilent model. Then look at each serial number listed for this group. If the
serial number of your power supply is prior to any of the serial number(s) listed, perform the change indicated in the
Change column. Note that, several changes can apply to your supply.
Serial Numbers
Change
Serial Numbers
Change
Model 6680A
Model 6683A
3318A   00111
3318A   00116
3502A   00126
3619A   00167
1, 2, 4, 5, 6
3
7
9
3349A   00104
3349A   00108
3514A   00123
3514A   00178
3619A   00203
6
3
7
8
9
Model 6681A
Model 6684A
3319A   00154
3336A   00164
3350A   00172
3350A   00183
3350A   00193
3506A   00218
3619A   00383
4
1
2, 6
5
3
7
9
3339A   00103
3405A   00108
3405A   00109
3514A   00118
3514A   00188
3514A   00257
5
3
6
7
8
9
Model 6682A
3349A   00104
3349A   00108
3514A   00123
3514A   00168
3624A   00193
6
3
7
8
9
CHANGE 1
In Table 5-14  A10 Control Board
DELETE
C548, 0.1uF 50V, Agilent P/N 0160-4835 in parallel with RT500.
CHANGE 2
In Table 5-7  A3 FET Board
DELETE
C228 & 231, 2000pF 100V P/N 0160-2301.
ADD
C228, 6800pF 100V P/N  0160-4904.
CHANGE
R247, from 75K P/N 0757-0462, to 42.2K 1% 1/8W P/N 0698-3450.
R249, from 13.1K P/N 0757-0444, to 7.5K 1% 1/8W P/N 0757-0440
R250, from 60.4K P/N 0698-3572 to 40.2K 1% 1/8W P/N 0698-3499.
R253, from 12.5K P/N 0698-6533 to 17.8K 1% 1/8W P/N 0698-3136.
C229,  from 0.022uF P/N 0160-4918 to 0.01uF 100V Ceramic P/N 0160-4832.