9.  The Uninstall 3M Molecular Detection System Uninstall dialog appears. You have the option to delete or to save the 
existing 3M Molecular Detection System database. The default is to save the database. To delete the database, click the 
Delete database checkbox.
10.  Click the 
[Uninstall] button to uninstall the software or click the [Cancel] button to cancel the uninstall action.
Uninstall Screen
Running a 3M Molecular Detection Instrument Self-Check
Before you begin
•  The 3M Molecular Detection Instrument Self-Check starts automatically upon power-on and can be started using the 
software. 3M recommends performing the Self-Check each day you use the 3M Molecular Detection Instrument.
•  The Self-Check includes the following tests:
 
– Firmware validity
 
– LED function
 
– Power distribution
 
– Heater function
 
– Photodiode function
 
– Memory integrity check
•  Do the following before starting a Self-Check to prevent failure:
 
–  Ensure that the pluggable heater is installed properly
 
– Remove the 3M Speed Loader Tray from the instrument
 
–  Close the instrument lid 
•  The Self-Check lasts approx. 2 min. During the Self-Check, the instrument status light cycles colors, the power light 
flashes red for a period, and then all lights turn off.
•  If you want to capture the start-up Self-Check result in the software, connect the instrument to the computer and log in 
to the software before turning on the instrument.