13.  Use a dry disposable towel to wipe the pluggable heater until it’s dry.
14.  Use a dry disposable towel to gently press down on each spring pin (1) on the pluggable heater.
15.  Set aside the pluggable heater in a spot where it can air dry for at least one hour.
16.  Use a disposable towel lightly dampened with 1-5% (v:v in water) household bleach solution to gently wipe the inner 
surfaces of the instrument. Wring out the disposable towel thoroughly so it is damp, but not dripping. 
17.  Use another disposable towel to repeat the procedure with DI water (distilled or deionized water).
18.  Use a dry disposable towel to wipe the interior surfaces of the instrument until they’re dry. Allow the interior surfaces of 
the instrument to air dry for at least 1 hour. Do not reconnect the power cable and the USB cable until you have allowed 
the instrument to air dry for at least 1 hour. 
19.  Connect the USB cable (3), connect the power cable (2) and turn on the power (1).
Packing the instrument for shipment to 3M for service
WARNING: always follow the cleaning and decontamination procedure prior to returning the instrument for service.
1.  Turn the 3M Molecular Detection System off using the on/off switch located on the instrument’s back panel.
2.  Remove the USB cable from the USB 2.0 port on the instrument’s back panel (5), and from your computer’s USB 2.0 port (6).