C. Entering the density of a liquid. 
  Two ways to set the density of a liquid are available in the function table, Liquid density 
input (ldin): by entering the water temperature or by entering the density directly. 
• Enter the water temperature (ldin 0) 
  The water temperature currently set (unit:
°C, 
factory setting : 25
°C) is displayed.   
 
  Use the following keys to change the value. 
RE-ZERO key   
To increase the temperature by one degree. (0-99
°C) 
MODE key   
To decrease the temperature by one degree. (0-99
°C) 
PRINT key   
To store the change, display    end    and return to the density mode 
(the condition of step1, (2) Selecting a parameter to set). 
CAL key   
To cancel the change and return to the density mode (the condition 
of step 1, (2) Selecting a parameter to set). 
The relation between the water temperature and density is shown below. 
 
 
 
 
 
 
 
 
 
• Enter the density directly (ldin 1)  
  The density currently set (unit : g / cm
3
, factory 
setting : 1.0000g / cm
3
) is displayed.   
   
  Use the following keys to change the value. 
RE-ZERO key   
To set the value of the digit selected. 
SAMPLE key 
To select the digit to change the value. 
PRINT key   
To store the change, display    end    and return to the density mode   
(the condition of step1, (2) Selecting a parameter to set). 
CAL key  
To cancel the change and return to the density mode  (the 
condition of step1, (2) Selecting a parameter to set). 
Note 
The range to set the density is 0.0000 to 1.9999 g / cm3   
 
(4) Outputting the density. 
After all the necessary parameters have been set, press 
the SAMPLE key to output the density calculated. 
Note 
The density is displayed with three decimal places.