A&D GP-100KS Benutzerhandbuch

Seite von 98
 
 12
3-6  Smart Range Function 
The GP-32K, GP-32KS and GP-102K are equipped with two ranges. The precision range has a 
higher resolution. The standard range has normal resolution. 
The range is switched automatically, depending on the value displayed. Pressing the RE-ZERO key 
allows weighing in the precision range, regardless of the tare value. (Smart range function) 
The range can be fixed to the standard range, by pressing the SAMPLE key. 
Note 
Once the range is switched to the standard range, it will not switch to the precision range 
automatically even when the displayed value becomes within the precision range value. Use 
the RE-ZERO or SAMPLE key to activate the precision range again. 
 
Below is a description of how the Smart Range actually functions, using the GP-102K   
(Precision range=61 kg
×1 g, Standard range=101 kg×10 g) 
 
1  Press the RE-ZERO key.   
  The balance will start weighing, using the 
precision range. 
2  Place a container, on the weighing pan, to be 
used as a tare. 
  The value displayed exceeds the precision 
range value. So, the range will be switched to 
the standard range. 
3  Press the RE-ZERO key.   
  The balance will be switched to the precision 
range. 
4  Place a sample on the pan.   
 
The value displayed is within the precision 
range value. So, the balance will perform a 
weighing, using the precision range. 
 
Precision range/standard range value 
 
GP-32K/GP-32KS GP-102K 
Precision range   
(after RE-ZERO is pressed) 
Up to 6.1009 kg 
Up to 61.009 kg 
Standard range 
6.101 to 31.008 kg 
61.01 to 101.08 kg