A&D GP-100KS Benutzerhandbuch

Seite von 98
 
 33
Operation 
1 Calibrate the balance using the internal mass. 
(one-touch calibration) Then, place an external weight 
and confirm the value to be corrected. 
  In the example, the value is to be corrected by 
0.6 gram in 20 kilograms. 
2  Press the ON:OFF key to turn off the display. 
3  While pressing and holding the PRINT key and the 
SAMPLE key, press the ON:OFF key. The balance 
displays   p5  .  
4 Press the PRINT key. Then the balance 
displays the function switches. 
 
 
 
 
  Set the function table switch and internal mass correction 
switch to “1” as shown above using the following keys. 
SAMPLE key 
To select the switch to change the 
value. The selected digit blinks. 
RE-ZERO key 
To change the parameter of the 
switch selected. 
5  Press the PRINT key to store the new setting. The 
balance returns to the weighing mode. 
6  Press and hold the SAMPLE key to enter the function 
table and release the key when    ba5fnc  is displayed. 
7  Press the SAMPLE key several times until  C5in   is 
displayed, then release the key. 
8 Press the PRINT key to enter the procedure for 
correcting the internal mass value. 
9  Correct the internal mass value using the following keys. 
RE-ZERO key 
To select the value. 
 
(-50 digits appear after +50 digits.) 
PRINT key 
To store the new value and display the 
next menu item of the function table. 
CAL 
key 
To cancel the correction and 
display the next menu item of the 
function table. 
10  Press the CAL key. The balance returns to the weighing mode. 
11  Press the CAL key to calibrate the balance using the internal mass. 
12  Place the external weight on the pan and confirm that the correction has been performed 
properly. In this example, confirm that the value displayed is within 
±2 digits of the correction 
reference value or 20 kilograms. If the value is not within 
±2 digits of the correction value, 
repeat the above procedure to correct it.