SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
30
Heavy Duty Cleaning Procedure
 WARNING
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare 
disinfectant. The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level different 
than that of Invacare disinfectant will result in an improper mixture of disinfectant and water. This may 
cause your system to be improperly disinfected - or in some instances, not disinfected at all. If using any 
disinfectant with a dilution ratio or viscosity level different than that of Invacare disinfectant or if 
switching from one type of disinfectant to another type, the disinfectant flowmeter must be calibrated 
before performing this procedure. Refer to Calibrating the Disinfectant Flow Meter on page 31.
ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting, cleaning and 
rinsing procedures. In case of eye or skin contact, follow procedures on disinfectant container.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The 
use of these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use 
cleaners containing iodine, bromine, Betadine or methylene blue. The use of these types of cleaners 
will cause the tub to stain. Never use cleaners containing bleach. Use of these types of cleaners will 
dry out the rubber seals and gaskets and the tub will not function properly.
Before starting the whirlpool pump, the water level must be completely above the whirlpool jets. 
Otherwise damage to the motor/pump will result.
In addition to disinfecting your tub after every use, you will need to perform this procedure approximately once 
every 80‐150 baths.
1.
Clean the entire surface of the tub to remove visible tissue 
residue and fluid.
2.
Rinse the inside surface of the tub with the shower/massage 
wand on the spray setting.
3.
Rotate the water temperature valve to adjust the water 
temperature to approximately 105
o
 F.
NOTE: This will maximize the cleaning solutionʹs ability to clean and 
disinfect the system.
4.
Close the drain.
5.
Make sure the whirlpool jets are open.
6.
Fill the tub one inch above the highest whirlpool jets.
7.
Add the Invacare Heavy Duty Cleaner to the water in the 
tub. Refer to the chart for the proper amount.
8.
Spray pre‐mixed Invacare Heavy Duty Cleaner on the 
remaining surfaces.
9.
Press the whirlpool on/off button to turn on the whirlpool 
pump and let the pump run for 60 seconds.
10. Turn off the pump and allow the heavy duty cleaner/water 
solution to remain in the tub for 10 minutes. 
11. During this ten‐minute waiting time, take a long handled 
brush and thoroughly scrub all interior surfaces of the tub.
12. Open the drain to allow the heavy duty/water solution to 
drain from the tub. 
13. Use the shower/massage wand on the pulse setting and 
direct the spray into both whirlpool jets to back flush until 
the water discharging from the whirlpool inlet is clear.
14. Rinse the tub thoroughly with the shower wand on the spray setting.
15. Check the tub for damage and to ensure all surfaces of the tub are clean and that no visible residue 
remains. If residue is found, repeat STEPS 13 and 14
TUB 
MODEL 
NUMBER
WATER REQUIRED 
TO COVER JETS 
(GALLONS)
INVACARE 
HEAVY DUTY 
CLEANER 
REQUIRED (OZ)*
2800
10
40
2001
32
128
3700
30
120
3750
30
120
3800
30
120
3600
30
120
3650
48
192
3600XL
48
192
2300
25
100
3300
30
120
4300
35
140
4300HL
35
140
5300
52
208
5300HL
52
208
6300
50
200
6300ADH
50
200
4400SS
48
192
5400SS
50
200
*NOTE: If using non‐Invacare heavy duty 
cleaner, refer to the instructions supplied with the 