TM
718A Air Ionizer Test Kit
When used with the 3M
TM
718 Static Sensor, the 3M
TM
718A
Air Ionizer Test Kit can provide periodic verification of air
ionizer performance.  The 718A Air Ionizer Test Kit consists
of a charge plate and a charger.  The system as a whole is
designed to meet the guidelines specified in ESD Association
Draft Standard D3.3.
The charge plate slides over the housing of the 718 Static
Sensor. Placing the 718 in the air flow of an air ionizer
allows 
for measurement of the ionizer’s offset balance.  The charger
can be used to deposit an electrostatic charge on the plate.
Measuring how long it takes the ionizer to neutralize this
static charge gives the ionizer discharge time.
Calibration and Certification
The 3M 718 Static Sensor can be ordered as calibrated 
and certified by a 3M ISO9002 Registered Metrology
Laboratory.  Calibration can be traced back to National
Institute of Standards and Technology (NIST) standards,
as specified in MIL-STD-45662.
3M 718A Air Ionizer Test Kit Specifications
Charge Plate Assembly
Per ESD Association Draft Standard D3.3
Charge Plate Bracket 
1" (H)  x  3.1" (W)  x  4.25" (L)
Dimensions
2.5 cm (H)  x  7.9 cm (W)  x  10.8 cm (L)
Charge Plate Area
3.25" (W)  x  1.25" (L)
8.3 cm (W)  x  3.2 cm (L)
Charge Plate Assembly 
Weight
2.5 oz (70 g)
Charger Dimensions
0.85" (H) x 2.4" (W) x 5.0" (L)
2.2 cm (H) x 6.1 cm (W) x 12.7 cm (L)
Charger Weight
6 oz. (170 g) with battery
Charger Power Requirements One 9 volt alkaline battery
Charger Output
1100 V minimum for positive or
negative voltage
Certifications
UL, C-UL, CE, CB-scheme, NOM
Electronic Handling & Protection Division  
6801 River Place Blvd. 
Austin, TX 78726-9000 
http://www.3M.com/ehpd
Recycled paper
40% Pre-consumer waste paper
10% Post-consumer waste paper
Litho in USA.
© 3M IPC 2001 98-0799-0493-8
For ordering information, technical information,
product information, you can reach us at:
Phone 1-800-328-1368
Fax 
1-800-828-9329    or    1-512-984-2210
3M 718A Air Ionizer Test Kit 
3
Important Notice
All statements, technical information, and recommendations related
to 3M’s products are based on information believed to be reliable,
but the accuracy or completeness is not guaranteed.  Before using
this product, you must evaluate it and determine if it is suitable for
your intended application. You assume all risks and liability
associated with such use.  Any statements related to the product
which are not contained in 3M’s current publications, or any
contrary statements contained on your purchase order shall have
no force or effect unless expressly agreed upon, in writing, by an
authorized officer of 3M.
Warranty; Limited Remedy; Limited Liability. 
This product will be free from defects in material and manufacture for
a period of 1 year from the date of purchase.  3M MAKES NO
OTHER WARRANTIES INCLUDING, BUT NOT LIMITED TO, ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR 
A PARTICULAR PURPOSE.
If this product is defective within the
warranty period stated above, your exclusive remedy shall be, at
3M’s option, to replace or repair the 3M product or refund the
purchase price of the 3M product. Except where prohibited by 
law, 3M will not be liable for any loss or damage arising from this
3M product, whether direct, indirect, special, incidental or
consequential regardless of the legal theory asserted.