Canon galvano gc-251 Benutzerhandbuch

Seite von 88
 
CANON    Digital Galvano Scanner System
 
KP-1SM Series + GC-251 
Users Manual 1.00 
 
 
 
  
49
Input the external trigger signal 
The external trigger signal is input from the digital I/O connector. 
See 5-1 ‘Connector Pin Arrangement) 
 
Connector pin arrangement 
Pin 
NO. 
Monitor Board 
Signal Name 
I/O Signal 
Description  Logic  Remarks 
B5 
10 Input 
External 
Trigger 
Signal  High:ON 
See Circuit below.
   
 
Connection Specifications 
   
3.3V TTL  Input 
 
 
 
 
Operation procedure 
(Raster Scan Parameter Settings) 
Set Parameter ID =26 
Raster Scan Time 
Set Parameter ID =27 
Raster Scan Duty Ratio 
   
 
Set Parameter ID =28 
Raster Scan Angle 
(Raster Scan Operation) 
   
 
Send Command ID=8    Data=6  Movement Start (Movement to the Initial position   
of raster scan) 
   
 
Send Command ID=23    Data=8  (One way scan)   
                         
      Data=9  (One coming and going) 
                                 
(Reference: Data =3 in case of continuous operation) 
        ・
Input the external trigger Signal 
    ・
Send Command ID=23  Data=0  Scan Stop 
 
Notes 
・ 
The external trigger input during raster scan move is ignored. 
・ 
10usec or less is uneven from the external trigger signal input at time until beginning to   
・ 
actually operate. 
 
FPGA 
3.3V-TTL 
R  :  100Ω 
Connector 
Select either