Emerson HAS60E-IM-HW Benutzerhandbuch

Seite von 72
Instruction Manual
HAS60E-IM-HW
June 2007
NGA 2000 CLD
3 - 17
Operation
Emerson Process Management GmbH & Co.OHG
Figure 3-23: Converter Test Setup
NO
X
   Mode
 
NO Mode
 
A.  TYPICAL TEST SETUP
 
B.  TYPICAL TEST RESULTS
 
TO SAMPLE INLET OF 
CLD ANALYZER MODULE
 
NO in
 
AIR 
ADDED
 
OZONATOR 
OFF
 
a-b
 
b
 
a
 
c-d
 
c
 
OZONATOR
 
ON
 
N
 
2
 
90
 
80
 
85
 
20
 
0
 
FLOWMETER 
FM2
 
METERING 
VALVE MV2
 
CONNECTOR C2
 
 
VAC
 
VARIABLE 
TRANSFORMER
 
OFF
T
1
OZONATOR 
UTILIZING 
ULTRAVIOLET 
LAMP
 
FLOWMETER 
FM1
 
METERING 
VALVE MV1
 
CONNECTOR C1
 
ZERO 
AIR
 
STANDARD GAS: 
NO IN N
2
 BACKGROUND
 
TO NO
X
 
MODE
 
d
 
ALL LINES AND FITTINGS STAINLESS STEEL OR TEFLON
 
ON
 
% Efficiency = 100
 
1+  a - b
 
c - d
 
CONNECTOR C3
 
FLOW APPROXIMATELY 
3 LITERS PER MINUTE
 
CAUTION: Externally limit sample 
flow rate to less than 2200 cc/min.
 
(            )