Emerson NGA2000 Benutzerhandbuch
3 Display and Keyboard
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 3
3.5
Important Functions of the Softkeys
DISPLAY
♦
Change from the single component display to the multi component display.
MEASURE
♦
Change from the main menu to the single component display.
♦
Change to the status report: Display of the most important parameters and
status informations of the CLD module or of the current channel.
status informations of the CLD module or of the current channel.
♦
If available, this command is always assigned to the F2 key.
CLD
F1
F2
F3
F4
F5
95.0 ppm
Basic Controls
Measurement range number:
4
Range and functional control:
LOCAL
Ranges with valid calibration:
4
Calibration status:
READY
If it won´t calibrate...
Status:
Status:
STANDBY
Ozonator status:
On
Ozonator power:
Enabled
Range upper limit:
250.0 ppm
Measurement mode:
NOx
HOME
NO/NOx
ZERO
SPAN
INFO
HOME
FACTORS
ZERO
SPAN
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Zero/span calibration
Measurement range number:
3
Span gas concentration:
100.0 ppm
Sample flow:
1100 ml/min
Raw measurement signal:
521590
Measurement gas:
NOx
NO/NOx toggle!
Result...
Calibration adjustment limits:
Calibration adjustment limits:
Disabled
Zero gas concentration:
0.0 ppm
Status:
READY
Line of Variables
Text Line
Function Line
Menu Line
Function Key Legend