Adam Equipment gbc-gfc series Benutzerhandbuch

Seite von 36
 
P a g e
 27  
                     
© Adam Equipment Company 2011
 
 
11.0  ERROR
 
CODES 
During the initial power-on testing or during operation, the scale may show an error 
message.  The meaning of the error messages is described below. 
If an error message is shown, repeat the step that caused the message, turning the 
balance on, carry out the calibration or other functions. If the error message is still 
shown contact your dealer for further support.   
 
ERROR CODE 
DESCRIPTION 
POSSIBLE CAUSES 
Err  1 
Time input error. 
Tried  to  set  an  illegal  time,  i.e. 
26hours  
Err  2 
Date input error 
Tried to set an illegal date, i.e. 36th 
day  
Err  4 
Initial  Zero  is  greater 
than  allowed  (typically 
4%  of  the  maximum 
capacity) when power is 
turned  on  or  when  the 
[Zero] key is pressed, 
Weight  is  on  the  pan  when  turning 
the scale on. 
Excessive  weight  on  the  pan  when 
zeroing the scale. 
Improper calibration of the scale. 
Damaged load cell. 
Damaged Electronics. 
Err 6 
A/D count is not correct 
when  turning  the  scale 
on. 
Platform is not installed. 
Damaged Load cell. 
Damaged Electronics. 
 
diSPEr 
Shows  when  a  number 
is  too  large  tot  fit  the 
display 
The  accumulated  totals  for  weight 
or  count  exceed  999999.    The  data 
on RS232 is still shown accurately