Adam Equipment 302666242 Benutzerhandbuch

Seite von 44
 
© Adam Equipment Company 2011 
Page  
18
 
 
6.3.4 Count pre-set or check-weighing 
 
Check-weighing  (or  Count  Pre-setting)  is  a  procedure  to  cause  an  alarm  to 
sound  when  the  net  weight  (or  the  number  of  items)  on  the  scale  meets  or 
exceeds a number stored in memory. The value to be stored is entered from 
the keyboard.     
 
SETTING OF PRESET LIMITS 
 
It is possible to set a high and low limit for either counting or weighing (using 
net  weight).    When  the  [Pst]  key  is  pressed  the  user  can  select  either 
counting or weighing and then set the lower and upper limit. 
 
For example: 
 
ACTION 
DISPLAYS 
Press [Pst] 
“CHECk  ”   “ nEt “    “         “ 
 if currently in the weighing mode 
 
Press 
[U.Wt/Units] 
“CHECk    ”      “  CoUnt  “        “                  “
    to  toggle  from  weighing  to 
counting 
Press [Print] 
“Hi  cnt”    “0.3  2  3  4”    “            ”
  The  current  high  count  limit  is 
displayed, press [CE] to clear and change if needed. 
If using net weight the display will show 
“Hi WEG”  
 
 
Press [Print] 
“lo  cnt”    “0.0  2  3  4”    “            ”
  The  current  low    count  limit  is 
displayed, press [CE] to clear and change if needed. 
If using net weight the display will show 
“Lo WEG”  
 
 
Press [Print] 
Return to weighing with limits set. 
 
 
 
  Clearing  of  either  or  both  the  limits  is  allowed.  Both  limits  cleared  will 
disable the Preset altogether. 
 
  If Preset Weight was selected the first display would show “
Hi WEG
” and 
Lo WEG
”.  
 
  The beeper will work as described in the Beeper Parameter.