Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection 
icon will light to notify you.
Body movement detection function
If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse 
wave icon appears in the measurement result display.
After 5 seconds, measurement resumes, and an attempt is made to complete measurement in one 
cycle.
Taking the Measurement in “Auscultation Mode”
Use a stethoscope to perform measurement.
When set to “ON”, SYS and DIA can be recorded using auscultation measurement.
To determine SYS and DIA during measurement, press the [MODE] button.
SYS is registered the first time and DIA is registered the second time you press the [MODE] button 
during deflation.
After DIA is determined, the cuff rapidly deflates and SYS and DIA are displayed as the measurement 
results.
The pulse rate does not appear in the measurement result display.
During deflation, the [
] button can be held down to re-inflate, or the [
] button can be held down to 
deflate faster.
Stored data that was measured in “Auscultation Mode” is displayed when the auscultation icon is lit.
Note:
The body movement detection function is disabled while “Auscultation Mode” is in use.
For Auscultation Mode settings, see page 16.