List of Error Codes
The alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority 
alarm occurs.
To clear an alarm, press any button.
Q If a low-priority alarm and a medium-priority alarm occur at the same time, the medium-priority alarm 
is displayed.
If the alarms are the same priority level, the error code of the alarm that occurred first is displayed.
However, to prevent battery consumption when powered by battery, a low battery error (E40 error) is 
always given priority.
Q Example: E2
SYSTEM
NIBP
Error 
code
Priority
Description
Points to check
E9
Medium
Internal hardware error
Contact OMRON HEALTHCARE.
Error 
code
Priority
Description
Points to check
E1
Low
The cuff tube is not connected
Firmly connect the cuff tube.
Air is leaking from the cuff.
Replace with an OMRON GS CUFF that 
does not leak.
E2
Low
Did not inflate properly because the arm 
or body moved during measurement.
Have the patient not move the arm or body, 
and repeat measurement.
Moved body or arm during 
measurement, or talked.
Have the patient not talk or move, and 
repeat measurement.
The cuff is not applied correctly.
Correctly apply the cuff.
The sleeve is rolled up and is 
compressing the arm.
Remove the garment and re-wrap the cuff.
Measurement time has exceeded 
specified time.
Specified time: 165 seconds
The measurement time exceeds the 
expected time, so the measurement was 
ended in order to avoid patient discomfort. 
There is a possibility that measurement is 
being repeated over and over due to air 
leaking from the cuff.