CHAPTER 11: PROGRAMMING GROUP 2 – METER CALIBRATION 
GE Industrial Systems                               5300P  Advanced Power Meter
 
67 
CHAPTER 11 
PROGRAMMING GROUP 2—METER CALIBRATION 
 
WARNING—READ THIS SECTION CAREFULLY BEFORE PROCEEDING: 
 
•  Any rescaling, such as a change in a transformer ratio, can be accomplished in GROUP 1. 
Calibration is not necessary. 
 
•  Meter calibration cannot be performed if the meter is installed for service. The sensing inputs must be 
connected to a power supply with variable voltage and separate current outputs. 
 
•  The calibration procedure requires highly accurate and stable input signals. Incorrect readings result 
from improper calibration procedures. If unsure, return unit to the factory for calibration. 
 
•  BEFORE calibrating any channel, make a note of its Full Scale Setting (see Chapter 10).  Set the Full 
Scale in accordance with Table 10-2 for calibration. Restore original Full Scale Setting when 
calibration is completed. 
 
 
11.1: Calibration Requirements 
 
Calibration on the EPM 5300P requires precise inputs of 120 Volts, 5 Amps, and 2.5 Amps. The EPM 
5300P -G model requires precise inputs of 300 Volts, 5 Amps, and 2.5 Amps. If this equipment is 
unavailable, contact us for assistance. 
 
GROUP AND FUNCTION NUMBER 
FUNCTION 
20. 
High End Calibration—Volts AN, BN, CN 
21. 
High End Calibration—Amps A, B, C 
22. 
Low End Calibration—Amps A, B, C 
2E. 
Exit Programming Group 2 
Table 11-1: Group 2 Programming Format 
 
NOTE:
  
The voltage and current scales are calibrated with the maximum range value 2000 for current, 
14.40 for voltage, 300 for option G.  The calibration value can be identical to the full scale value in 
GROUP 1, for better accuracy than the specification.  However, if a full scale is ever changed to a higher 
value than the calibrated value, the value may be less accurate.