TRouBLESHooTING
If any abnormality arises during use, please check the following points:
Note:  If the unit still does not work, contact HoMedics Consumer 
Relations. Under no circumstance should you disassemble 
or attempt to repair the unit by yourself. Contact informa-
tion for HoMedics Consumer Relations Department can be 
found on the warranty page (See page 34).
32
31
SYMPTOMS
POSSIBLE 
CAUSES
CORRECTION
Unit does not 
turn on when the 
START/STOP 
button is pushed.
Batteries have run 
down.
Replace them 
with two new AAA 
alkaline batteries.
Battery polarities 
have been positioned 
incorrectly.
Re-insert the 
batteries in the 
correct positions.
EE measurement 
error symbol shown 
on display or the 
blood pressure 
value is displayed 
excessively low (or 
high).
Cuff has been placed 
incorrectly.
Wrap the cuff 
proerly so that it is 
positioned correctly.
Did you talk or 
move during 
measurement?
Measure again. 
keep wrist 
steady during 
measurement.
Shaking of the wrist 
with the cuff on.
E1 error symbol
shown on display
Air circuit 
abnormality.
Measure again.
E2 error symbol 
shown on display
Inflation pressure 
exceeding 300 mmHg.
Switch the unit off, 
then measure again.
E3 error symbol 
shown on display
Error determining 
measurement data.
Measure again.
PoTENTIAL foR ELECTRoMAGNETIC 
INTERfERENCE 
To avoid inaccurate results caused by 
electromagnetic interference between 
electrical and electronic equipment, do 
not use the device near a cell phone or 
microwave oven.