4.4.1 
Formaldehyde Disinfection Preparation 
WARNING: 
Always wear rubber gloves, apron and facemask.  Use proper ventilation during 
disinfection and rinse-out.  A face respirator with an organic cartridge to remove 
formaldehyde fumes may be worn.  DO NOT BREATHE FUMES OR ALLOW 
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES. Follow all 
Material Safety Data Sheet (MSDS) precautions and OSHA Standards. 
WARNING: 
Label machine with appropriate warning signs such as "Do Not Use / Contains 
Disinfectant" to prevent use of the RO until it is properly rinsed. 
CAUTION: 
New or replacement thin film membranes must be rinsed to drain with normal pressures 
for a minimum of six (6) hours before first exposure to formaldehyde, or irreversible 
damage may occur.  
Disinfecting Materials 
Description 
GE Osmonics Part Number 
Disinfection/Cleaning Kit  
12214 
Formaldehyde, 37% Solution 
Not Provided by GE Osmonics 
Formaldehyde Test Kit 
Not Provided by GE Osmonics 
“DO NOT USE” Warning Labels   Not Provided by GE Osmonics 
Cleaning/Disinfect Log Sheets 
Reference Tech Notes 143, 144, 145 
Operation Log Sheet 
Reference Tech Note 142 
Body Protective Wear 
Not Provided by GE Osmonics 
1238339b – 16Oct03 
4 - 16 
Routine Maintenance