GE Z-8000 Benutzerhandbuch

Seite von 128
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Z Series RO System by GE Osmonics 
CAUTION: 
‘RENALIN®’ is corrosive to most metals other than stainless steel. Any brass 
components that will come into contact with the ‘RENALIN®’ must be replaced with 
suitable stainless or plastic materials prior to attempting disinfection. ‘RENALIN®’ type 
products contain oxidizing agents and are potentially damaging to thin film (polyamide) 
membranes.  It is imperative that all the precautions and instructions are followed 
correctly.  Failure to comply with these instructions, which incorporate instructions from 
the membrane manufacturer, will jeopardize the system performance and the warranty. 
Renalin®, Minncare™, Peracidin™ and others are trade names for compounds based 
on Peracetic Acid and Hydrogen Peroxide compounds, and are common in dialysis 
facilities.  The products referred to in this procedure typically have 20% hydrogen 
peroxide and 4% peroxacetic acid before dilution.  Other products are available with 
these same chemicals in different ratios and are not acceptable for use in the Z-Series 
RO.  Actril™ and other compounds with different combinations of these chemicals 
must not be used on polyamide (thin-film) membranes.
  The name, “Renalin®” will 
be used in this manual to refer to acceptable products. 
Disinfecting Materials 
Description 
GE Osmonics Part Number 
Disinfection/Cleaning Kit  
12214 
Renalin®, 1 quart 
Not Provided by GE Osmonics 
Renalin® Test Kit 
Not Provided by GE Osmonics 
“DO NOT USE” Warning Labels   Not Provided by GE Osmonics 
Cleaning/Disinfect Log Sheets 
Reference Tech Notes 143, 144, 145 
Operation Log Sheet 
Reference Tech Note 142 
Body Protective Wear 
Not Provided by GE Osmonics
1238339b – 16Oct03 
4 - 21 
Routine Maintenance