Omron hem-907 Benutzerhandbuch

Seite von 26
13
Select the arm cuff according to the arm circumference of the patient.
Arm circumference
Name of the arm cuff
17 - 22 cm
22 - 32 cm
32 - 42 cm
Small size adult arm cuff (option)
Adult arm cuff
Large size adult arm cuff (option)
Be sure to use the arm cuff suitable for the size of the patient’s arm.
If an arm cuff not suitable for the size of the arm is used, blood pressure 
may not be measured correctly.
Securely connect the air tube.
If you connect the attached 1m air tube, the air tube can be used with the 
entire length of 1.2 m.
Prepare the patient to be ready to wrap the arm cuff.
Wrap the arm cuff over the right upper arm’s skin as a rule.
If the arm cuff is wrapped over a thick cloth or by rolling up the jacket or shirt, blood pressure may not be measured 
correctly.
Place the right hand of 
the patient with the palm 
of hand facing upward.
4.
Place the arm cuff over the patient’s 
arm by matching the mark indicating 
the artery position 
to the brachial artery.
5.
RANGE
MIN
MAX
IND
EX
ART.
22~32
cm (9"
~13")
REF
HEM
-CR19
ART.
Wrap the arm cuff snugly using both hands and securely fasten it with the Velcro tape. At this time, the lower 
edge of the arm cuff must be placed 1 to 2 cm above the inner side of elbow joint.
If the arm cuff is not wrapped snugly, patient may feel numbness on his/her arm, or blood pressure may not be 
measured correctly.
The arm cuff wrapped diagonally along the shape of the arm does not affect the measurement value.
MIN
ART.
RANGE
MIN
INDEX
ART.
22~32
cm(9"
~13")
REF
HEM
-CR19
RANGE
1-2 cm
1.
2.
3.
6.
How to Apply the Arm Cuff
RANGE
MIN
MAX
IND
EX
ART.
22~3
2cm
(9"~1
3")
REF
HEM
-CR19
IND
EX
ART.
RANGE
MIN
MAX
32~42
cm(13
"~17"
)
REF HE
M-CL1
9
ART.
RANGE
MIN
MAX
IND
EX
REF H
EM-CS
19
17~22
cm (7"
~ 9")
Warning
Do not apply the arm cuff during the intravenous 
injection or transfusion.
Intravenous injection or transfusion is impossible.
Requests from OMRON
Do not use the unit to the patient using a pump oxy-
genator.
Do not in ate the unit without applying the arm cuff.
Do not use the damaged cuff.
Cuff side
Air tube 
side
ART.