Omron i-c10 Benutzerhandbuch

Seite von 48
32
3. Using the Unit
3.5
Using the Memory Function
The unit is designed to store the blood pressure and the pulse rate in 
the memory for two people (user A and user B) every time a 
measurement is completed.
The unit automatically stores up to 84 sets of measurement values 
(blood pressure and pulse rate) for each user (A and B). When 84 sets 
of measurement values are stored, the oldest record is deleted to save 
the most recent values. The unit also stores 8 weeks of morning 
averages and 8 weeks of evening averages for each user (A and B).
Average Function
The unit calculates the average reading based on the three most 
recent sets of measurement values taken within 10 minutes of the most 
recent reading.
Note: If two sets of measurement values are stored in the memory for 
the 10 minutes period, the average is based on the two sets of 
measurement values. If one set of measurement values is 
stored, this is displayed as the average.
To Display the Measurement Values
1.
Select your user ID (A or B).
2.
Press the MEMORY button (
).
The average value is shown and the average symbol (
appears above the memory symbol (
) on the display.