ELECTRODE CONDITIONING
AND MAINTENANCE
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will
disappear when rinsed with water.
During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode
Reference
Filling Hole
Reference
Filling Hole
Fill Hole
Screw
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Glass
Bulb
Reference
Junction
Reference
Junction
Reference
Junction
Platinum or
Gold tip
Platinum or
Gold tip
Reference
Wire
Reference
Wire
Sensitive
Wire
Glass Body
pH Electrode
Plastic Body
pH Electrode
Plastic Body
ORP Electrode
Glass Body
ORP Electrode
Reference
Wire
In the event that measurements are per-
formed with solutions containing sulfides or
proteins, the cleaning of the diaphragm of
the reference electrode must be performed
(see page 21, "Cleaning Procedure").
In order to have a correct functioning of the
ORP electrode, immerse it into HI 7020 and
measure the response; the obtained value
should be within 200 and 275 mV.
After this functional test, it is suggested to
wash the electrode thoroughly with water
and proceed to the oxidizing or reducing
pre-treatment before taking measurements.
When not in use, the electrode tip should be
kept moist and far from any type of me-
chanical stress which might cause damage.
For this reason , the use of the protective
cap supplied with the electrode is advised.