Hanna Instruments hi 3832 Benutzerhandbuch

Seite von 1
 
Instruction Manual
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instructions carefully before using the
chemical test kit. It will provide you with the necessary
information for correct use of the kit.
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:
• 1 color comparator cube;
• Reagent 1 (30 mL);
• Reagent 2 (20 mL).
Note: Any damaged or defective item must be returned in
its original packing materials.
 
HI 3832
Iodine Test Kit
ISTR3832R1      06/00     PRINTED IN ITALY
Iodine may be used as a disinfectant in drinking and swim-
ming pool waters. Unlike chlorinated pools, water treated
with iodine decrease eye irritation among swimmers and
provides a disinfectant more stable to adverse conditions.
However, iodine levels in the water must be frequently
checked to maximize its given purpose.
The Hanna Iodine Test Kit employs a color cube to deter-
mine the iodine content in water. The kit is compact and is
practical for field use. No chlorine or bromine can be present
in the water sample for this test to work properly.
 Range
0 to 2.5 mg/L (ppm) Iodine
 Smallest Increment
0.5 mg/L (ppm) Iodine
 Analysis Method
Colorimetric
 Sample Size
25 mL (average)
 Number of Tests
50 (average)
 Case Dimensions
220x145x55 mm (8.7x5.7x2.1")
 Shipping Weight
180 g (6.7 oz.)
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USE
CHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTION
The iodine is measured by a colorimetric method. The
sample is initially treated with a phosphate buffer to a pH
of approximately 6.3. The addition of DPD (N,N-diethyl-p-
phenylenediamine) is immediately oxidized by iodine
producing a reddish color. The color intensity of the solution
determines the iodine concentration.
REFERENCES
REFERENCES
REFERENCES
REFERENCES
REFERENCES
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Add 7 drops of Reagent 1 and 3 drops of Reagent 2 in
the color comparator cube and mix by carefully swirling
the cube in tight circles.
• Fill the color cube with water
sample to the 25 mL mark.
• Cap and invert several time to mix.
• Determine which color matches the solution
in the cube and record the results as mg/L
(or ppm) iodine.
INSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONS
Standard Methods for the Examination of Water and Waste-
water, 16
th
 Edition, 1985.
The chemicals contained in this kit may be hazardous if
improperly handled. Read Health and Safety Data Sheet
before performing this test.
HEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETY