Report No.: RM920221R04A
18
Issued: Mar. 14, 2003
Reference No.: RM920221R04
5.4.5 TEST SETUP
P
S
P
S
P
S
P
S
EUT
Vertical 
coupling plane
0.5mm Isolation Support
ESD 
Generator
Nearest 
Wall
Horizontal 
coupling plane
Reference Ground Plane
0.1m
1m
470k
X4
80cm
For the actual test configuration, please refer to the related item – Photographs of the
Test Configuration.
NOTE:
TABLE-TOP EQUIPMENT
The configuration consisted of a wooden table 0.8 meters high standing on the Ground Reference
Plane. The GRP consisted of a sheet of aluminum at least 0.25mm thick, and 2.5 meters square
connected to the protective grounding system. A Horizontal Coupling Plane (1.6m x 0.8m) was placed
on the table and attached to the GRP by means of a cable with 940k
Ω total impedance. The equipment
under test, was installed in a representative system as described in section 7 of EN 61000-4-2:1995,
and its cables were placed on the HCP and isolated by an insulating support of 0.5mm thickness. A
distance of 1-meter minimum was provided between the EUT and the walls of the laboratory and any
other metallic structure.