Instruction Manual 
 
CON 11/110 
 
24
5.4 
Using Manual Ranging Function 
Your meter automatically selects the range in which your readings appear. The 
manual ranging function lets you select the specific range you want to work in: 
Meter Range Symbol 
Conductivity Range 
TDS Range 
r1 
0 to 19.99 µS/cm, 
0.00 to 9.99 ppm 
r2 
20.0 to 199.9µS/cm  
10.0 to 99.9 ppm 
r3 
200 to 1999 µS/cm; 
100 to 999 ppm 
r4 
2.00 to 19.99 mS/cm, 
1.00 to 9.99 ppt 
r5 
20.0 to 199.9 mS/cm
 
10.0 to 99.9 ppt 
5.4.1  Selecting manual range 
1.  To select the desired measuring range, 
press the RANGE key while in 
measurement mode. A momentary display 
of either “r1” to “r5” will flash indicating the 
range you are in. The range selected will 
appear on the display and the “MEAS” 
indicator blinks indicating that you are in 
manual range. 
 
2.  Press the range key again (if needed) until 
desired range is selected. To re-select the 
auto-ranging function, repeatedly press the 
range key until the “MEAS” indicator appears 
without blinking.  
NOTES 
If the value of the solution you are measuring is higher than the range selected, “Or” 
will appear on the primary display. Press RANGE until the correct range is selected.  
The meter resets to the Auto-ranging function once it is 
turned off. You will have to reset the manual ranging function 
each time you turn the meter off.  
MEAS
READY
READY
1413
ATC
°C
22.3
µS
MEAS
1413
ATC
°C
22.3
µS
RANGE
MEAS Blinking
READY
MEAS
ATC
°C
22.3
µS
Figure 19: Selecting Manual 
Range 
Figure 20: Out of 
Range