The following table provides a guideline to enable diagnosis of possible problems 
indicated by the messages generated by the meter. The table also provides possible 
solutions to the problems encountered. 
 
ERROR MESSAGE 
INDICATES 
POSSIBLE CAUSE 
CORRECTIVE 
ACTION 
ERR. Annunciator 
Wrong keypad input 
Wrong input in 
selected mode.  
Release key. Select 
valid operations 
depending on mode.  
Secondary Display 
Toggles Continuously 
between Calibration 
Standard Values 
Calibration error  
Incorrect calibration 
solution used or 
contamination of  
calibration solution 
occurring in automatic 
calibration mode 
Check if the right 
calibration solution 
was selected or use 
fresh calibration 
solution. 
Battery icon lights up 
Low Battery 
Battery power is low 
Replace batteries 
with fresh ones as 
soon as possible.