eutech-instruments cyberscan con 11 conductivitytds meter Benutzerhandbuch

Seite von 80
Instruction Manual 
 
CON 11/110 
 
71
15 SPECIFICATIONS 
 
SPECIFICATIONS 
CON 11 
CON 110 
Conductivity Range 
0 to 19.99 µS/cm, 
20.0 to 199.9µS/cm,  
200 to 1999 µS/cm; 
2.00 to 19.99 mS/cm, 
20.0 to 199.9 mS/cm 
0 to 19.99 µS/cm, 
20.0 to 199.9µS/cm,  
200 to 1999 µS/cm; 
2.00 to 19.99 mS/cm, 
20.0 to 199.9 mS/cm 
TDS Range 
0.00 to 9.99 ppm 
10.0 to 99.9 ppm 
100 to 999 ppm; 
1.00 to 9.99 ppt 
10.0 to 99.9 ppt 
0.00 to 9.99 ppm 
10.0 to 99.9 ppm 
100 to 999 ppm; 
1.00 to 9.99 ppt 
10.0 to 99.9 ppt 
Resolution 
0.05 % Full Scale 
0.05 % Full Scale 
Accuracy 
±1% F.S. +1 digit 
±1% F.S. +1 digit 
Temperature Range 
-10.0 to 110.0
°C 
-10.0 to 110.0
°C 
 
 
(14.0 to 230 
°F) 
Temperature 
Resolution 
°C 
0.1 
°C 0.1 
°C 
Temperature 
Resolution 
°F 
Not available 
0.1 
°F(from 14.0°F to 199.9°F) & 
1.0
°F(from 200°F to 230°F) 
Temperature Accuracy ± 0.5 
°C ± 
0.5 
°C (±0.9 °F) 
Cell Constant, K 
0.1, 1.0, 10.0 (Selectable) 0.1, 
1.0, 
10.0 (Selectable) 
Temperature 
Compensation 
Automatic / Manual (from 0 to 
100°C) 
Automatic / Manual (from 0 to 
100°C) 
Temperature 
Coefficient 
0.0 to 10.0% / ° C 
0.0 to 10.0% / ° C 
Normalisation 
Temperature 
20.0 or 25.0°C 
20.0 or 25.0°C 
TDS Factor 
0.40 to 1.00 
0.40 to 1.00 
Calibration Points 
(Automatic) 
84.0, 1413µs/cm; 12.88, 111.8 
mS/cm  
84.0, 1413µs/cm; 12.88, 111.8 
mS/cm  
Calibration Points 
(Manual) 
User defined standard / 1 per 
range up to 5 points. 
 
User defined standard / 1 per 
range up to 5 points. 
 
Single/Multi point 
calibration 
Selectable  
Selectable  
Calibration Range 
10% to full scale  
10% to full scale  
Auto-ranging 
Yes (manual select available ) 
Yes (manual select available ) 
HOLD Function 
Yes 
Yes 
Memory 
50 Data Sets 
100 Data Sets 
Auto Off 
Selectable 
Selectable 
   Auto Off Time 
20 minutes after last key 
operation 
20 minutes after last key 
operation