Instruction Manual 
 
CON 400/410 
- 10 - 
4.4 Conductivity 
Calibration 
 
The CON 400 and CON 410 meters are capable of up 
to 5-point conductivity calibration at one point per 
conductivity range (0.00 - 19.99 
µS; 0.0 - 199.9 µS; 0 - 
1999 
µS; 0.00 - 19.99 mS; 0.0 - 199.9 mS).  
 
All new calibration data will over-ride existing stored 
calibration data for each measuring range calibrated.  
 
Calibrating for Conductivity: 
1. 
If necessary, press the MODE key to select 
conductivity mode.   
2. 
Rinse the probe thoroughly with de-ionized (DI) 
water or a rinse solution, then rinse with a small 
amount of calibration standard.  
3. 
With the yellow probe guard attached, dip the 
probe into the calibration standard. Immerse the 
probe tip beyond the upper steel band. Stir the 
probe gently to create a homogeneous sample. 
Allow time for the reading to stabilizse.  
See Figure 7.  
4. Press CAL/MEAS to enter conductivity 
calibration mode.  The CAL indicator will appear 
in the upper right corner of the display.  
5. Press 
the 
MI/
S or MR/T key to change the 
value on the primary display to match the value 
of the calibration standard referenced to your 
normalisation temperature (usually 25
°C).  
6. Press ENTER to confirm calibration value.  The 
meter returns to the MEAS (measurement) mode.  
7. 
Repeat steps 1 to 6 for other measuring ranges.  
 
See figure 6.  
 
 
 
 
 
 
 
Figure 6 - Conductivity Calibration