8  TEMPERATURE CALIBRATION 
IMPORTANT: When Calibration mode is entered, controller automatically goes into 
a “HOLD” mode where the 4-20 mA output freezes and relays are de-activated (if it 
was in an activated condition). Upon return to measurement mode, both 4-20mA 
output and relay activities resume, depending on settings. 
 
This controller features selectable Automatic Temperature Compensation (ATC) or 
Manual Temperature Compensation (MTC). 
ATC: ATC mode requires a Pt 100 temperature element. ATC automatically 
compensates for temperature fluctuations. You can offset your ATC temperature 
reading by ±10
o
C. 
Important: If there is no temperature element wired to the controller and ATC is 
selected on, the screen will flash in pH or ORP mode, and you will see an error 
message (OR) in temperature mode. 
MTC: MTC lets you select a specific value at which temperature will be 
compensated. You can select a manual temperature value from – 10 to 110
o
C. 
Factory default is 25.0
o
C. 
See Setup program P3.3 for instructions on ATC or MTC selection. 
To offset temperature: 
1. 
Press MODE to select 
o
C mode. The display shows MTC or ATC, then the 
temperature. 
2. 
Press CAL / ENTER key to directly access the offset mode when the unit is 
in temperature measurement mode. 
NOTE: Instead of pressing the CAL/ENTER key, you can also enter the offset 
mode using the password code procedure as explained in Section 5. 
3. 
The screen will flash the current 
o
C reading. 
4. 
For ATC: Determine the temperature of your solution with an accurate meter 
(such as the Temp 5). Press the ▲ or ▼ 
keys to offset the 
o
C value on the controller 
display to match the value of the solution you 
are measuring. 
For MTC: Press the ▲ or ▼ keys to offset 
the 
o
C value on the controller display to 
match the desired value. 
5. 
Press ENTER. The display flashes “DONE” 
and return to the temperature measurement 
mode.
 
 
22.5
25.0
done
ENTER