9.7  Viewing pH Electrode Data – P5.0 
In pH mode, Program 5 has three “view only” options that lets you check the 
electrode parameters for diagnostic purposes. 
P5.1: view pH electrode offset 
P5.2: view pH electrode slope 
P5.3: view temperature probe offset (ATC on only) 
 
These parameters will change each time you recalibrate the controller. 
P5.0
elec
Slpe
P5.1
P5.2
P5.3
p.ofs
t.ofs
0.0
5 .
92
5
.
0
 
Press ▲ and ▼ keys together (ESCAPE) at anytime, to leave the Setup mode 
and return to Measurement mode. 
 
P5.0: Viewing pH electrode data 
1. 
Key in the password “022” as per procedure in Section 5. 
2. 
Press  ▲ key four times. The screen will scroll P5.0 and ELEC. Press 
ENTER. 
3. 
The screen will scroll P5.1, then P.OFS, then pH electrode offset in mV at 
the pH 6.86 / 7.00 calibration point. 
4. 
Press ENTER. The screen will display P5.2, then SLPE, then the electrode 
slope in mV. 
5. 
MTC on: press ENTER to return to general Setup mode. 
ATC on: go to next step. 
6. 
The screen will scroll P5.3, then T.OFS, then the temperature offset value. 
7. 
Press ENTER to return to general Setup mode.  
Press ▲ and ▼ keys together (ESCAPE) to return to measurement mode.