eutech-instruments ecoscan do 6 do meter Benutzerhandbuch

Seite von 76
Instruction Manual 
DO 6 
 
 
7
 
3.4 
 Dissolved Oxygen Probe Information 
The DO 6 palm top meter is supplied with a 
Dissolved Oxygen Probe that works on galvanic 
principle, that is, it does not require any 
polarising voltage from your meter. Rather it will 
generate a millivolt signal proportional to the 
amount of oxygen in the solution.  
Your DO Probe has a twin cable, one with a 
BNC connector for the DO measurement input, 
and the other with a phono jack plug for the 
temperature measurement input. Its sensing 
area consists of a cathode, anode, and an inner 
electrolyte which is separated from your sample 
solution by an oxygen permeable membrane 
pre-membraned to the probe’s detachable cap. 
Re conditioning of your probe is made easy with 
the introduction of this specially designed 
detachable pre-membraned cap. 
This light weight probe (93X233912 /35642-50) 
has a built-in temperature sensor for Automatic 
Temperature Compensation (ATC) with 12mm 
diameter epoxy body housing and 16 mm 
diameter Noryl detachable cap. Its compact 
sensing area reduces air entrapment which 
makes it easy to obtain fast, accurate and 
stable readings. Simply stir the probe in the 
solution being measured. Shaking will aid to 
remove bubbles if needed before taking a 
reading. Also, always ensure to remove water 
drops from the membrane when calibrating in 
air. 
Proper use of probe is essential to ensure that 
the optimum measurement is taken in a short 
Figure 5: 12 mm Galvanic DO 
probe  (93X233912 / 35642-50)
 
Detachable 
Pre-membraned Cap 
(Probe Sensing Area)
16 mm
12 mm