Operating Manual  
alpha DO 1000 
 
12
CALIBRATION MODE  
You can reach the Calibration mode directly from the 
Measurement mode by pressing the CAL key and 
entering the Calibration security code. You can also 
reach the Calibration mode from the Advanced Setup 
mode.  
 
6.1 Dissolved 
Oxygen 
Calibration  
If the units of measure is in mg/l (default), then the 
calibration is also carried out in mg/l. The Dissolved 
Oxygen Controller allows a one- or two-point calibration.  
 
1. Enter Calibration mode. While in the 
Measurement mode, push the CAL key and 
scroll to Calibration code “11”.  Push the 
ENTER key again. The upper and lower display 
reads “CAL dO”. 
2.  Press the ENTER key. Use the ▲ or ▼ keys to 
select 1- or 2-calibration points.  
3. For One-Point Calibration, only Hi-value 
calibration is possible. The probe is allowed to 
equilibrate in a solution containing 8.24 mg/l of 
dissolved oxygen. The “CAL” indicator flashes 
at the top, right-hand corner. When reading is 
stable, the “READY” annunciator comes on. 
Use the ▲ or ▼ keys to adjust the displayed 
value to the correct value. Press ENTER key to 
accept the value and the controller displays the 
status of the probe.  
4. For Two-Point Calibration, calibration is 
carried out at the Lo-value first, usually zero 
oxygen solution. Immerse the probe in the 
solution (low levels of dissolved oxygen) and 
allow it to stabilise. When the reading is stable, 
the “READY” annunciator comes on. Use the ▲ 
or  ▼ keys to adjust the displayed value to the 
correct value. Press the ENTER key to accept 
the value. 
NOTE: If the displayed value is > ±0.4 mg/l, the controller will not 
accept the zero calibration and retains the previous calibration 
values.  