eutech-instruments waterproof c401 colorimeter Benutzerhandbuch
Calibration: Chlorine, Free & Total
C401/301/201/102/101
25
10. Select another clean vial, un-screw the cap and fill it with 10 ml of
the prepared Chlorine standard (fill up to the line on the vial).
11. Select a DPD Free sachet, or a DPD Total sachet, depending on
the selected test. Open the sachet and tip the contents into the vial
containing the calibration standard.
containing the calibration standard.
12. Replace the cap, ensuring it is screwed on finger tight to
the vial. Invert the vial repeatedly for approximately 20
seconds, until the powder has dissolved.
seconds, until the powder has dissolved.
NOTE: a small amount of the powder may remain un-dissolved in the
vial, but this will not affect the measurement.
vial, but this will not affect the measurement.
13. Insert the vial fully into the sample well of the meter, whilst again
aligning the ▼ mark on the vial with the ▲ mark on the meter.
14. For Free Chlorine calibration, press the READ/ENTER key within
two minutes from the time the sachet was added to the vial. As for
Total Chlorine calibration, wait for two minutes before pressing the
READ/ENTER key and this has to be done within five minutes from
the time the sachet was added to the vial.
Total Chlorine calibration, wait for two minutes before pressing the
READ/ENTER key and this has to be done within five minutes from
the time the sachet was added to the vial.
15. The meter will display [--Rd--] whilst the calibration standard is
being measured and then displays [SAVE] to prompt for calibration
acceptance and saving of calibration data.
acceptance and saving of calibration data.
Free
Chlorine
Sample
4.5 ppm
Chlorine
Sample
4.5 ppm
ppm
READ
ENTER
CAL
16. Press the READ/ENTER key. The meter now accept the calibration
value and momentarily display [HoLD] and CAL annunciator before
returning to the beginning of the calibration as in step 4.
returning to the beginning of the calibration as in step 4.
17. To continue with another calibration, press the MODE key to select
the desired calibration mode.
18. To exit calibration mode, press ZERO key to return to the
measurement mode.
MODE
READ
ENTER
Press Mode to select
another calibration test
another calibration test
ZERO
Press ZERO to return
to measurement mode
to measurement mode