dhi m connector bloc for rpm4 pn 401929 installation instructions Benutzerhandbuch

Seite von 1
 
RPM4 and //m are trademarks of DH Instruments, Inc.
 
Document 560073 
041122 
DH Instruments, Inc. 
4765 East Beautiful Lane 
Phoenix  AZ  85044-5318 
USA 
 
 
Tel 602.431.9100
Fax 602.431.9559
dhi@dhinstruments.com
www.dhinstruments.com
 
 
 
//m™ 
Connector Bloc for RPM4™ 
P/N 401929 
Installation Instructions 
//m CONNECTOR BLOC, P/N 402168, 
INCLUDES: 
QTY 
PART NO 
DESCRIPTION 
402169 
//m connector bloc 
2 101006-Z 
M3 
screw 
102464-Z 
M3 split washer 
103345 
Adapter, 1/8 in. M O-seal x 1/8 in. swage 
1 560073 
Instruction 
sheet 
INTRODUCTION 
The RPM4 //m connector bloc is an accessory for 
RPM4 parallel measurement mode operation. It 
supplies single TEST(+) and TEST(-) ports to assure a 
simple, leak free connection to both of the RPM4’s Q-
RPTs in parallel. 
The //m connector bloc can only be used with RPM4s 
that have two Q-RPTs of A7M or lower. Maximum 
working pressure is 7 MPa (1 000 psi).  
INSTALLATION INSTRUCTIONS 
  Install the four 1/8 in. M O-seal x 1/8 in. swage 
adapters on the RPM4 TEST(+) and TEST(-) ports 
(2).  Remove the nuts and ferrules from each one of 
the adapters. 
  Remove the two Philips head screws from the 
extreme right and left hand corners of the RPM4 Hi 
and Lo 
 
Q-RPT modules (1, 3). 
  Remove the cover of the //m connector bloc by 
removing the two top (5) and two lateral (4) button 
head socket screws. 
  Connect the four 1/8 swage F connections of the 
//m connector (6) to the four 1/8 swage M 
adapters on the RPM4 TEST(+)  and  TEST(-) 
ports. 
 
Attach the //m connector bloc (7) to the RPM4 
using the two M3 screws supplied with the kit.  
Install the screws in the lower right hand corners 
of the Q-RPT modules (1, 3). 
 
Leak check all connections and reinstall the //m 
connector bloc cover. 
 
Figure 1.  //m connector bloc 
 
Figure 2.  RPM4 rear panel Q-RPTs 
(2 Q-RPTS of A7M or lower)