7) Editing Logger/Repeater Settings
a) Once the system has been setup existing Loggers & Repeaters can be
modified. All fields except for Unit ID can be modified. There are two
options for changing Logger/Repeater setup:
i) Wireless Change: Allows changes to all fields except for Unit ID and
options for changing Logger/Repeater setup:
i) Wireless Change: Allows changes to all fields except for Unit ID and
Push to Start
(1) Logger must be actively communicating
(1) Logger must be actively communicating
– no lost
communication errors
(2) From the Equipment Status Screen right click on the Logger and
select edit to open Edit Logger window.
(3) Modify field. More than one field can be changed at a time.
(4) Click on SAVE CHANGES AND START. A window will pop up
(4) Click on SAVE CHANGES AND START. A window will pop up
indicating that the change has been sent via Wireless mode.
Leave the Logger in its current location until the wireless change is
complete.
Leave the Logger in its current location until the wireless change is
complete.
(5) The unit can not be edited and the Edit option will remain grayed
out until the changes have been completed. If after 25 minutes the
Edit option is still grey, bring the unit to the PC and edit the unit via
USB.
Edit option is still grey, bring the unit to the PC and edit the unit via
USB.
(6) NOTE:
(a) When changing a Logger/Repeater wirelessly, the change can
not be sent until the Logger wakes up to send data and may
require more than 1 attempt to send all changes. Depending on
the sample rate, this might take up to 25 minutes.
require more than 1 attempt to send all changes. Depending on
the sample rate, this might take up to 25 minutes.
(b) If the change needs to take place immediately use the USB to
modify setup.
ii) USB Change: Use this option for faster changes and in the event
that the Logger has lost communications
(1) Connect the Logger to the PC via the USB cable.
(2) The Edit Logger window will open.
(3) Modify field(s). More than one field can be changed at a time.
(4) Click on SAVE CHANGES AND START. The Updating Logger
(1) Connect the Logger to the PC via the USB cable.
(2) The Edit Logger window will open.
(3) Modify field(s). More than one field can be changed at a time.
(4) Click on SAVE CHANGES AND START. The Updating Logger
window will open.
(5) Once the update is complete, unplug the USB and place the