System Component Overview:
Host PC:
Host PC:
PC where the WiZARD2 Software and Receiver are installed and Loggers
and Repeaters are setup.
Only one WiZARD2 System can operate on one PC at any time.
To view logged data from more than one PC, the host PC and non host
PCs must be connected to a network and the non host PCs must have
WiZARD2 Network Software installed. See section 11 below for more
information.
WiZARD2 Network Software installed. See section 11 below for more
information.
This PC must be on 24/7 in order for the WiZARD2 system to operate
properly.
It is recommended that the host PC be rebooted occasionally to keep
things running smoothly. There will not be any lost data while rebooting
Individuals installing and setting up the WiZARD2 system must have full
administrative rights on the host PC.
USB Connection: Best when connected to primary USB. Will not operate
on an unpowered USB hub.
WiZARD2 Software:
Software is included with the Receiver and is required to setup, manage
and view all Loggers, Repeaters and saved logged data.
Stores all logged data points for viewing via graphs and tables in an
encrypted database.
Must run 24/7 on host PC in order for WiZARD2 system to operate
properly.
Receiver (WR220/225):
WR220 features WiZARD2 Software. WR225 features WiZARD2
SECURE 21CFR11 Compliant Software.
The Receiver forwards data packets from the Loggers to the WiZARD2
database.
Must always be on and connected to the PC via a USB cable. It is
recommended that it be powered via the AC adapter and have Backup
Batteries installed.
Batteries installed.
Receivers do not require calibration.
Logger (WT200/220/240/WH220/225/245):
Take temperature or temperature/humidity sample points at regular
intervals and send those readings to the Receiver at regular transmission
intervals.
intervals.
Loggers are setup via USB at the host PC.
Have a practical operating range of up to 200 ft from Receiver. Any metal
infrastructure or other wireless devices will significantly reduce operating
range.
range.
Will save logged data when communications with the Receiver have been
interrupted. Once communications are restored, saved logged data will be
sent.
sent.
Loggers should be calibrated every 6 to 12 months.