VIVIX-S 1012N User Manual 
6.4 
Calibrating by Loading the Calibration Data 
 
 
Vieworks provides the calibration data stored in the detector. Although the Offset 
calibration is performed automatically, you need to download both Defect and Gain 
calibration data since their calibration cannot be performed in the detector automatically.   
6.4.1  Preparing Calibration Data 
Among the three types of calibration data stored in the detector, download the Defact Map and Gain 
calibration data to local HDD. 
 
 
It is not necessary to download the Offset calibration data since the data is used in the detector. 
File 
Description 
Offset 
Detector Offset data 
Gain 
Calibration data for image sensitivity 
Defect Map 
Defect calibration map data 
6.4.2  Loading Defect Calibration Data 
 
 
1  Select Defect Calibration from the Step list. 
2  Click Download Defect Map button to assign the path of local HDD, and download the Defect Map 
calibration data. 
3  Click Load Defect Map button to load the calibration data. 
 
 
 
The calibration menu status is displayed as Done when the Defect calibration data is 
loaded completely.