If the remaining of battery is under 30% or at the 1
st
 level, the system warns low battery 
and the detector will be turned off automatically if the battery is being consumed for a 
specific period of time. Therfore, it is recommended to change the battery when a 
warning message or indicator is displayed. 
7.1.3  Wireless Communication Diagnosis 
 
 
In case of using the detector with wireless communication way, make sure to check the 
status of wireless communication before starting to use. If the status of wireless 
communication is bad, the speed of acquiring images will be very slow or failed to acquire 
images. Try to check the surrounding wireless communication status not to occur 
communication interference. If wireless communication module in the detector has 
problems, consult the sales representative in Vieworks or a relevant engineer. 
 
Check the Connection Status 
User can check the connection status of wired or wireless detector through VXvue or VIVIX SDK
 
Check the Strength of Wireless Communication Signal 
The signal strength of wireless communication is provided as 5 stages or percentage (%), user can check it 
through VXvue (Vieworks Viewer) and VIVIX SDK
 
Level 
Level 
Link Quality 
Meaning 
Very Good 
66 ~ 70 
The communication is running smoothly, and it 
ensures performace of the image acquisition. 
Good 
56 ~ 65 
Normal 
41 ~ 55 
The communication status is normal, but it does not 
ensure performace of the image acquisition. 
Bad 
31 ~ 40 
The communication status can become unstable. 
Very Bad 
1 ~ 30 
Impossible to communicate normally. 
Unknown 
The communication is disconnected. 
 
 
You can check the wireless communication state from the Diagnosis dialogue in VIVIX Setup program.