VW40-152-009
 
VIVIX-S 1012N User Manual 
4.2.2  Booting up SCU 
 
 
1  Turn on the power switch at the front side of SCU. 
2  Check if the power switch of SCU lights up green. 
3  Check if the status indicator is blinking in green after the power switch of SCU is permitted. 
4  Check if the status indicator of SCU is blinking in green and turns to light up in the same color. 
 
 
 
When the power switch lights up green, it means that the power is applied normally. 
 
When the status indicator is blinking in green, the SCU is booting. 
 
If the status indicator lights up green, it indicates that the SCU is being operated normally 
after booting up. 
 
If the status indicator lights up blue, SCU’s preparation for communication is completed. 
4.2.3  Booting up the Detector 
 
 
The procedures in this chapter are explained according to the default setting status of a 
detector (factory reset) and the use of wireless connection. 
 
Refer to <7.4 Product Initialization> for the detailed information about the default setting 
status of a detector. 
 
1  Attach a charged battery pack to the detector. 
2  Press the power button of the detector for 1 second until the power LED is turned on. 
3  When the power LED is on, check if it is lit green. 
4  Check if the active LED is lit orange on the detector. 
 
 
 
If the power LED lights up green, the DC power is permitted normally. 
 
If the active LED lights up orange, the detector is compeleted to boot up normally.