Vieworks Co. Ltd. FXRD-1012NAW Benutzerhandbuch

Seite von 151
 
 
Rev.1.0
 
Page 88 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
WNetwork 
Sets AP (Access Point) information of wireless communication from the detector. 
Item 
Description 
SSID 
Configures wireless network ID AP from the detector. 
Key 
Configures wireless network key value of AP from the detector. 
Wireless Only 
Configures the wireless communication method of the detector. 
On 
The detector is operated in a wired way when a tether interface cable is 
connected. 
Off 
The detector is operated in a wireless way when a tether interface cable is not 
connected. 
 
AP 
Checks and sets AP (Access Point) of detector information. 
Item 
Description 
AP On / Off 
Decides whether to run the detector on AP (Access Point) mode. 
On 
Turns on the detector AP 
Off 
Turns off the detector AP 
Frequency 
Frequency channel of wireless network 
2.4  ㎓ 
Uses 2.4  ㎓  Frequency. (Up to 13 channels) 
5  ㎓ 
Uses 5  ㎓  Frequency. (Up to 8 channels) 
Country 
Country code of wireless network (KR, US, EU, JP, CN) 
Band 
Wireless network bandwidth 
20  ㎒ 
Basic Frequency Band 
40  ㎒ 
Expands bandwidth through channel bonding. 
Channel 
Wireless network channel 
SSID 
Wireless network ID 
Key 
Wireless network key value (Applied to the password only.) 
 
 
 
13 channels can be used in 2.4  ㎓  Frequency. 
 
8 channels can be used in 5  ㎓  Frequency 
 
The number of serviceable channels is different according to the configured country. 
 
Channel bonding is used for enhancing transmission speed. However, the speed may be 
slowed down due to the interference of surrounding channels, even if the channels have 
been bonded. 
 
Channel items (+/-) will be activated in case of using 40㎒  frequency bandwidth. You can 
set whether to bond channels with the above or below one. 
 
SSID and Key values of the detector for communicating with SCU wirelessly (WNetwork) 
should be set as the same.