270 MHz Antenna 
 
Model 50270s 
 
Signal Position 
Place the antenna on the ground and use the Automatic Signal Position selection. The system 
will servo and place the direct coupling pulse at the top of the time range window. 
To test that you have the correct position, raise the antenna off the ground and you will observe 
on your system that the antenna transmit pulse will separate from the reflection from the ground. 
The higher that you raise the antenna, the further apart will be the two pulses. 
To assure that the direct coupling pulse (time zero) is recorded the user should place the signal 
Position servo in the manual mode. The signal should then be moved down in the time range 
window until the entire surface pulse is visible and there is some ‘dead time’ or flat trace visible 
above the direct coupling pulse in the time range window. 
Gain Check 
The surface pulse should be about 2/3 the width of the screen. If it is greater, reduce the Gains 
manually. If the signal appears too small you can manually increase the Gains, but the first gain 
point should never exceed 10dB, the last gain point should not exceed 65dB. 
Specifications 
Center frequency: 270 MHz 
Pulse duration: 3.6 ns 
Depth of penetration: 0-25 ft depending on dielectric permittivity  
Size of sensor: 17.5
×17.5×7.5 inches (44.5×44.5×19cm) 
Weight of sensor: 19 lbs (8.6 kg)