Symptom
Cause
Action
Probe fails to power up 
(no LEDs illuminated or 
fails to indicate current 
probe settings).
Exhausted batteries.
Change batteries.
Unsuitable batteries.
Change batteries.
Batteries inserted incorrectly.
Check battery insertion.
Batteries removed for too short a 
time and probe has not reset.
Remove batteries for a minimum 
of 5 seconds.
Probe fails to switch on.
Exhausted batteries.
Change batteries.
Batteries inserted incorrectly.
Check battery insertion.
Probe out of range.
Check position of RMI or RMI-Q, 
see operating envelope.
No RMI or RMI-Q 'start/stop' 
signal (radio on mode only).
Check RMI or RMI-Q for green 
start LED.
Incorrect spin speed (spin switch 
on only).
Check spin speed and duration.
Malfunctioning shank switch 
(shank switch mode only).
Check switch operation.
Incorrect switch on method 
configured.
Check configuration and alter as 
required.
Incorrect multiple probe mode 
setting configured.
Check configuration and alter as 
required.
RMP60 in hibernation mode 
(radio on mode only).
Ensure probe is in range and wait 
up to 30 seconds, then resend 
switch on signal. 
Check position of RMI or RMI-Q, 
see operating envelope.