Agfa Gevaert N V 5243 Benutzerhandbuch

Seite von 3
 
 
 
 
Confidential 
Page 1 of 3 
06/07/2011 
F:\agbmi05\Personal\PCBs\HE\Cirrus\CirrusRFIDreaderFCC\CirrusRFIDreaderRegulatoryInetgrationInstr
uctions.doc 
a
a
a
a
b
b
b
b
 
Medical Imaging
 
RDE
 
 
 
From:
 
Jan Vercammen (6233) – MI/TSP-HW/Electronics
 
 
Date: 
July 6, 2011 
Concerning: Regulatory and hardware instructions for integration of Cirrus 
RFID reader into host applications (concerns FCC ID: HPL5243)  
This document describes the regulatory elements and measures for the integration of 
a full modular short-range RFID reader into host applications. The internal company 
reference of the RFID reader (or reader for short) is A800626
 
The reader is a small printed circuit board with local shielding, local power supply regulation and 
filtering, here is a short summary of the measures taken to control (spurious) emissions: 
  The reader transmitter (chip) has a local PCB shield 
  The reader module power supply is 5V+/-10% and using a local low dropout regulator a 
3.3V+/-2% is used on board for all electric circuits, there is also sufficient bulk and high-
frequency decoupling 
  A common-mode (CM) choke is used on the 5V power supply as a means to limit 
unwanted CM conducted- and radiated emissions 
  The tuning circuit uses low tolerance components (1% or 2%) to limit circuit variations 
 
 
The full-modular approval allows the integration of the reader into various applications 
without the need for radio type testing of the application, however, the user needs to 
take a number of measures and comply with regulatory requirements