Troubleshooting
21
You replaced the batteries but still 
received the “Battery low” message.
If you just printed diagnostic reports, let the 
HF PARTNER system recharge for 30 to 60 
minutes. If you still hear this message after 
letting the batteries recharge, call Guidant 
Technical Services. See “Contacting 
Guidant” on page 5.
The HF PARTNER device cannot 
communicate with an implanted 
device.
Move the HF PARTNER device around 
gently until communication is made with 
the implanted device (see Figure 3 on 
page 7), or tr
y repositioning the 
HF PARTNER device (such as turning it 
over). 
Verify the implanted device is a 
compatible Guidant device.
Voice messages cannot be heard.
Turn the speaker On using the steps on 
page 12.
Voice messages are no longer in the 
language of your choice.
The selected language may have been 
changed. Reset the HF PARTNER device to 
the language of your choice. See “Language 
selection” on page 11.
The report won’t print
Change the distance between the 
HF PARTNER device and printer. 
Optimal performance is 3-15 feet.
Check that the printer adapter is turned 
On (look for flashing lights when the 
transmission occurs).
Check the printer for warning lights.
Pair with printer again.
System status report will not print 
properly during printer pairing 
If a new printer is being used, the printing of 
the test page may interfere with the pairing. 
Clear the printer and start the pairing 
process again. 
Problem
Possible Solution